1 - 10 of 40 Search Results for 60 micron water filter
|
|
6.
Innova300C_fredmotofred_preinstallation
(PAGE 14 OF 32)
|
|
|
... / MotoFreD Ion Laser 6 Table 3. Utility Requirements POWER REQUIREMENTS 208 VAC ±10% 1, 3-phase with ground, no neutral, WYE configuration, 50 or 60 Hz. Phase balance <3% between phases. Maximum current consumption (typical): 50 Amps per phase at 208 VAC 2 COOLING WATER REQUIREMENTS Flow rate, minimum ...
|
|
|
|
7.
Innova_322C_Preinstallation
(PAGE 10 OF 30)
|
|
... 2 Table 1. Utility Requirements PARAMETER REQUIREMENT POWER REQUIREMENTS Power requirements 215-229VAC1,2, 3-phase with ground, no neutral, 50 or 60 Hz, WYE configuration. Phase balance <3% between phases. Maximum current consumed (typical) 50 Amps per phase at 215 VAC 3 COOLING WATER REQUIREMENTS ...
Related Searches: innova 90 manual
|
|
|
|
8.
Innova_300C_preinstallation
(PAGE 15 OF 30)
|
|
... 208 VAC ±10%1 (All models except Innova 302C) 208 VAC +10%1, -7.5% (Innova 302C) 3-phase with ground, no neutral, WYE configuration, 50 or 60 Hz. Phase balance <3% between phases. MAXIMUM CURRENT CONSUMED (TYPICAL) MAXIMUM HEAT LOAD PRODUCED Innova 301C/304C 45 Amps per phase at 208 VAC 2 15 kW ...
Related Searches: innova 200
|
|
|
|
9.
Innova90C_preinstallation
(PAGE 15 OF 32)
|
|
... Table 3. Utility Requirements, Argon and Krypton LINE VOLTAGE REQUIREMENTS 208 VAC +10%1, 3-phase with ground, no neutral, WYE configuration, 50 or 60 Hz. Phase balance <3% between phases. MAXIMUM CURRENT CONSUMED (TYPICAL) MAXIMUM HEAT LOAD PRODUCED Innova 90C-3 40 Amps per phase at 208 VAC 2 13 ...
Related Searches: innova 90c
|
|
|
|
10.
Laser_Cooling_Water_Guidelines
(PAGE 10 OF 34)
|
|
Innova Ion Laser Systems 4 Table 1. Innova Water Flow Specifications LASER SYSTEM MAXIMUM STATIC PRESSURE1 FLOW RATE MIN MAX PRESSURE DELTA RANGE6 CV7 EQUIVALENT ORIFICE DIAMETER8 KPa2 (psig3) lpm4 (gpm5) KPa (psig) mm (in) Sabre, Sabre R, Sabre Krypton Sabre SkyLight, Sabre PureLight Sabre MotoFreD ...
Related Searches: innova 300 | innova 70c | innova 70
|
|