00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

…three phase 480 VAC in WYE configuration, with no neutral and the yellow/green wire to building ground. There is no neutral connection. The Innova Sabre MotoFreD ion laser system will operate with either 60 Hz or 50 Hz line frequency. Electrical power must meet the specifications in Table 1. The…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Preface v Preface This manual provides preinstallation instructions for the Innova Sabre MotoFreD ion laser. Please refer Chapter Three, Laser Safety, in the operator's manual which describes laser safety features and precautions. If you are unfamiliar with ion lasers in general, refer to Chapter…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Preface v Preface This manual provides preinstallation instructions for the Innova Sabre MotoFreD ion laser. Please refer Chapter Three, Laser Safety, in the operator's manual which describes laser safety features and precautions. If you are unfamiliar with ion lasers in general, refer to Chapter…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Preface v Preface This manual provides preinstallation instructions for the Innova Sabre MotoFreD ion laser. Please refer Chapter Three, Laser Safety, in the operator's manual which describes laser safety features and precautions. If you are unfamiliar with ion lasers in general, refer to Chapter…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Innova Sabre MotoFreD Ion Laser Operator's Manual 4 It is important to note that the minimum temperature delta between the plant side of the Innova Sabre MotoFreD ion laser's heat exchanger and the laser side is 5°C (9°F). To determine the minimum pressure requirements of the facility water, the…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Innova Sabre MotoFreD Ion Laser Operator's Manual 4 It is important to note that the minimum temperature delta between the plant side of the Innova Sabre MotoFreD ion laser's heat exchanger and the laser side is 5°C (9°F). To determine the minimum pressure requirements of the facility water, the…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Innova Sabre MotoFreD Ion Laser Operator's Manual 4 It is important to note that the minimum temperature delta between the plant side of the Innova Sabre MotoFreD ion laser's heat exchanger and the laser side is 5°C (9°F). To determine the minimum pressure requirements of the facility water, the…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

…Cooling Water The Innova Sabre MotoFreD ion laser requires a flow of water to cool the laser head and electrical components inside the power supply. Because the properties of the cooling water are critically important to the performance of the laser, the Innova Sabre MotoFreD ion laser is delivered…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Innova Sabre MotoFreD Ion Laser Operator's Manual 8 Do not use de-ionized water in the internal cooling loop. Water that is high in resistivity (>2 M-cm) will slowly dissolve metal components in the cooling path. Dissolved metals, as with dissolved salts and minerals, can result in these materials…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

…relation. Altitude The maximum operational altitude for the Innova Sabre MotoFreD ion laser is 2,000 meter maximum. Site Preparation To assist in planning the work space, the dimensions of the components constituting the Innova Sabre MotoFreD ion laser system are listed in Table 3. Enough space…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

…Cooling Water The Innova Sabre MotoFreD ion laser requires a flow of water to cool the laser head and electrical components inside the power supply. Because the properties of the cooling water are critically important to the performance of the laser, the Innova Sabre MotoFreD ion laser is delivered…

00 SabreMFreD Preinstall.book

00 SabreMFreD Preinstall.book

Innova Sabre MotoFreD Ion Laser Operator's Manual 8 Do not use de-ionized water in the internal cooling loop. Water that is high in resistivity (>2 M-cm) will slowly dissolve metal components in the cooling path. Dissolved metals, as with dissolved salts and minerals, can result in these materials…

Powered by Site Search powered by SLI Systems

Call 1-800-227-8840


Outside the U.S. 1-408-764-4983

Coherent has locations across the globe that are available to provide support for any product, service or inquiry.

Visit our Contact Page to connect with any of our global sites.


Corporate Headquarters
Coherent Inc.
5100 Patrick Henry Drive
Santa Clara, CA 95054 USA

email: tech.sales@coherent.com

Send Us a Message

Please Read Before Continuing

Risk factors: Except for the historical information contained here, many of the matters discussed in this Web site are forward-looking statements, based on expectations at the time they were made, that involve risks and uncertainties that could cause our results to differ materially from those expressed or implied by such statements. These risks are detailed in the “Factors That May Affect Future Results” section of our latest 10-K or 10-Q filing. Coherent assumes no obligation to update these forward-looking statements.


I have read the terms and conditions