2.
Coherent Inc. CO2 RF Warning Lights On
|
|
|
…operation. If VSWR is still illuminated try replacing the RF cable (system off, facility power disabled), if fault doesn't clear see step 5. * 3rd Check - Forward warning indicator on. The most likely cause for this warning indicator is the RF amp. If the Reflected fault light is on, set system to ~ 10%…
|
|
|
|
|
7.
Coherent Inc. CO2 No Laser Output Beam â DIAMOND OEM
|
|
|
…On, troubleshoot as necessary. * Verify that RF cable connections at the laser head and the RF amp, are okay, not burnt verify that the RF cable is not damaged. Replace cable if necessary. * If DC voltage is okay. Disconnect 25 pin connector from RF amp. Click here for illustration. Using Oscilloscope…
|
|
|
|
|
8.
Coherent Inc. CO2 Laser System Doesnât Turn On â OEM Version
|
|
…power supply from the RF amp. Verify that 48 V is at the output of the DC power supply while disconnected. If not replace the DC. * Reconnect the RF amp to the DC power supply. Verify that 48 V is at the input of the RF amp. Click here for illustration. If not replace the RF amp. * Turn off the 48 V supply…