1SDG_HSD.QXD (Page 1)  (PAGE 2 OF 2)
…Benelux +31 (30) 280 6060
France +33 (1) 6985 5145
Germany +49 (6071) 9680
Italy +39 (02) 34 530 214
UK +44 (1353) 658 800
SDG
System Specifications
Trigger Input TTL-compatible, 1-50 kHz
RF Sync 10-100 MHz, >100 mV peak-to-peak
Outputs 1 fixed, 3 adjustable…
Related Searches: sdg

2SDG_HSD.QXD (Page 1)  (PAGE 1 OF 2)
…input trigger is precisely synchronized to
the RF signal by high-speed electronics.
The SDG produces four delayed trigger
signals, one fixed and three user adjustable in 250 ps steps. Two of the adjustable
trigger signals are typically used to trigger
Pockels cell drivers. The…
Related Searches: synchronization and delay generator | sdg | high speed driver | regenerative

3UCLD_Catalog.book  (PAGE 26 OF 37)
…beam paths, Demonstration of alignment optimization, System optimization, Software control and instructions, Maintenance procedures, Optics cleaning, SDG timing, Diode calibration,
Turn on/off procedures, Discussion of available/typical diagnostics, Troubleshooting scenarios.
Prerequisites: A…

4pdf document  Libra Preinstallation Manual Best match in this PDF  |  All matches in this PDF
SDG) The SDG controls the precise timing of the RA Pockels cells. It contains a high-speed power supply for the Pockels cells as well as a bandwidth detector (BWD) circuit, which serves as an interlock to protect the laser from operation at narrow bandwidth. Power Supplies The Libra system includes two
Related Searches: eye damage | sdg | legend elite user manual | evolution manual

5pdf document  UCLD Catalog Customer 4-12 Best match in this PDF  |  All matches in this PDF
…Excimer · Ion · Pulsed · Ultrafast Established 1966 He adquarters - Santa Clara, California U.S. A. Coherent Training Catalog - Description of Courses ii This document is copyrighted with all rights reserved. Under copyright laws, this document may not be copied in whole or in part, or reproduced in any…
Related Searches: coherent silhouette ii | halogen filter | optical coherence | ls670

6Libra PreInstallation.book  (PAGE 4 OF 35)
….... 3-5
3-3. Vitesse Power Supply Dimensions .................................................................................. 3-6
3-4. SDG Elite Dimensions..................................................................................................... 3-7
LIST OF TABLES
2…

7Libra PreInstallation.book  (PAGE 3 OF 35)
…Optical Component Labels ............................................................................... 2-2
Synchronization and Delay Generator (SDG) ........................................................... 2-6
Power Supplies ...............................................................…

8Libra PreInstallation.book  (PAGE 32 OF 35)
…system.
An external interlock connector is provided on the SDG Elite rear
panel. In the event of an interlock fault during normal operation, the
Pockels cells are disabled which terminates Libra output. Refer to
the SDG Elite Operator's manual for more details.
Alternatively, the…
Related Searches: sdg

9Libra PreInstallation.book  (PAGE 24 OF 35)
Libra Pre-Installation Manual
2 - 6
Synchronization
and Delay
Generator (SDG)
The SDG controls the precise timing of the RA Pockels cells. It
contains a high-speed power supply for the Pockels cells as well as
a bandwidth detector (BWD) circuit, which serves as an interlock…
Related Searches: libra manual | sdg

10Libra PreInstallation.book  (PAGE 26 OF 35)
…V 50/60 HZ (AMPS) 220 ± 20 V 50 HZ (AMPS)
Vitesse Power Supply 15 10
Evolution Power Supply 15 10
Water Chiller[1]
15 10
SDG Elite 1 0.5
Laptop computer 1.5 1
[1] All system components will operate on either 110 V or 220 V line voltage EXCEPT the water chiller.