Preinstallation Manual VerdiTM V-2/V-5/V-6 Diode-Pumped Lasers 5100 Patrick Henry Drive Santa Clara, CA 95054 Verdi V-2/V-5/V-6 Laser Preinstallation Manual ii This document is copyrighted with all rights reserved. Under the copyright laws, this document may not be copied in whole or in part or reproduced ... Related Searches:verdi v5 manual | mira | verdi-v2
... .6 lb.) Umbilical Length 3.8 m (12.5 ft.) Power Supply Dimensions 235 x 505 x 254 mm (9.3 x 19.9 x 10.0 in.) Power Supply Weight 31.5 kg (69.3 lb.) Chiller Dimensions 203 x 274 x 378 mm (8.0 x 10.8 x 14.9 in.) Chiller Weight 11.3 kg (24.9 lb.) MRU (Miniature Recirculator Unit) Dimensions 478 x 491 x ... Related Searches:azure 266 | 266nm cw laser | field max top
Preinstallation Manual VerdiTM V-8/V-10 Diode-Pumped Lasers 5100 Patrick Henry Drive Santa Clara, CA 95054 Verdi V-8/V-10 Laser Preinstallation Manual ii This document is copyrighted with all rights reserved. Under the copyright laws, this document may not be copied in whole or in part or reproduced ... Related Searches:verdi manual | verdi laser manual | verdi 10 | water chiller
... experience Supporting critical environments Highest level of post-sale customer care Long-term customer relationships Clogged Water Flow For Avia, Verdi and Vitesse lasers – Ensure distilled water is used in the chiller. Do not use deionized water, or hard water. Deionized water is corrosive. Hard ... Related Searches:water chiller | chiller | verdi laser manual
... 300W typical Frequency 50 to 60 Hz Laser Head Cooling Requirements7 Conductively-cooled (or optional heat sink, with or without a closed-cycle chiller) Power Supply Cooling Requirements Air-cooled Range of Operating Temperature 15°C to 35°C (59°F to 95°F) Weights Laser Head 10 kg (22.0 lbs.) Power Supply ... Related Searches:verdi 5 | verdi-v5 | v5 | verdi 5w
... Noise1 ±2.0% ±2.0% ±2.0% ±2.0% Stability2 ±2.0% ±2.0% ±2.0% ±2.0% M2 <2.0 <2.0 <2.0 <2.0 *Specifications based on output of one OPA, 10W Verdi pumping a RegA 9000/9050, and either Mira/Verdi or Vitesse seeding. Same specifications apply for two simultaneously pumped OPAs at 100 kHz. 1 Peak ... Related Searches:verdi price list
... 300W typical Frequency 50 to 60 Hz Laser Head Cooling Requirements7 Conductively-cooled (or optional heat sink, with or without a closed-cycle chiller) Power Supply Cooling Requirements Air-cooled Range of Operating Temperature 15°C to 35°C (59°F to 95°F) Weights Laser Head 10 kg (22.0 lbs.) Power Supply ... Related Searches:verdi v6 | verdi 5w | verdi-v5 | verdi-v2
Verdi V-8/V-10 Laser Preinstallation Manual 5 6. Proceed to the paragraph titled, "Installing the Power Supply". The riser/heat sink clamps and chiller connection will be installed during the final installation procedures. Installing the Power Supply 1. Move the power supply to an accessible ... Related Searches:verdi v10 manual
... post-sale customer care Long-term customer relationships For Vitesse lasers – Verify the baseplate is 25 ± 1ºCelsius. Optimum alignment of the Verdi pump beam into the ultrafast head can only be achieved if the baseplate temperature is 25 ± 1ºCelsius. Adjust the chiller temperature as necessary ... Related Searches:baseplate verdi
Verdi V-8/V-10 Laser Preinstallation Manual 6 Each customer must inspect the power cord and install the proper connector if necessary. The connector should be installed in a properly grounded outlet with a maximum of 16 Ampere service @ 110 V and 11 Ampere service @ 220 V for proper overcurrent and ... Related Searches:verdi laser manual | cooling