1 - 10 of 25 Search Results for verdi-v2

1pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 6 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual vi
Related Searches: verdi v5 manual

2pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 7 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 1 Introduction The purpose of this document is to assist the user in establishing a suitable location and operating environment for optimum Verdi V-2/V-5/V-6 performance. Receiving and Inspection Inspect shipping containers for signs of rough handling ...

3pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 8 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 2 1. Riser/heat sink installation hardware 2. Operator's manual 3. AC power cord 4. Power supply key 5. Installation kit case 6. Interlock-disassembled 7. Fuses and instructions 8. FAP-I installation hardware, including shorting clips, end caps, etc. 9 ...
Related Searches: verdi laser manual | verdi v5 manual

4pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 9 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 3 The laser head is shipped with a protective plastic plate installed over the baseplate. Leave the plastic plate in place until laser head is ready for installation. 1. Remove the packing material from the riser/heat sink (Figure 1-2). 2. Insert two dowel ...

5pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 10 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 4 1. Dowel holes (2) 2. Cooling water inlet and outlet (if included) 3. Laser head clamp screw holes (4) 4. Optical table clamp mounts (4) Figure 1-2. Riser/Heatsink 1. Laser head 2. Protective plastic cover retaining screws (3) 3. Protective material ...
Related Searches: verdi laser manual

6pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 11 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 5 6. Proceed to the paragraph titled, "Installing the Power Supply". The riser/heat sink clamps and chiller connection will be installed during the final installation procedures. Installing the Power Supply 1. Move the power supply to an accessible ...

7pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 12 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 6 Each customer must inspect the power cord and install the proper connector if necessary. The connector should be installed in a properly grounded outlet with a maximum of 10 Ampere service @ 110 V and 11 Ampere service @ 220 V for proper overcurrent ...
Related Searches: water chiller

8pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 13 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 7 7. Perform the procedure in paragraph titled, "Turn-on (Cold-Start)" in the Operator's Manual. External Interlock The system will not operate with the interlock open. An interlock connector is located on the power supply rear panel. The interlock ...

9pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 14 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 8 5. The interlock circuit in Verdi operates from a +/-12 VDC circuit. Its current is limited to around 1 mA. For these ratings, a "dry-circuit" type switch will give the most reliable operation. Figure 1-6. Interconnect Diagram Table 1-1. Utility ...

10pdf document  Verdi_V2-V5-V6_Preinstallation  (PAGE 15 OF 18)
Verdi V-2/V-5/V-6 Laser Preinstallation Manual 9 Dimensions Table 1-2. Environmental Requirements PARAMETER REQUIREMENT Operating temperature 15 to 35C (59 to 95F) Relative humidity 5 to 95% (non condensing) Altitude Sea level to 10,000 feet Figure 1-7. Laser Head Dimensions